Title of article :
Development of a pixellated Ir-TES for a large format TES array
Author/Authors :
Ohno، نويسنده , , M. and Takahashi، نويسنده , , H. and Fukuda، نويسنده , , D. and Kunieda، نويسنده , , Y. and Nakazawa، نويسنده , , M. and Ohkubo، نويسنده , , M. and Hirayama، نويسنده , , F. and Ohtani، نويسنده , , Vincent C. and Shimizu، نويسنده , , H.M.، نويسنده ,
Pages :
3
From page :
414
To page :
416
Abstract :
To realize imaging X-ray spectroscopy with high-energy resolution, we proposed a new pixellated array of Ir-TES. In this type of device, the bias point of each pixel is slightly modified. Therefore the analysis of the pulse shape parameter, such as the pulse height, the rise time and the fall time of a signal pulse, can identify the incident pixel position. We have obtained 13 eV (FWHM) at 3 keV energy resolution and 80 μm position resolution for 10 pixel devices. At this moment we could successfully operate a 20 pixel device.
Keywords :
Microcalorimeter , Pixellted Ir-TES , Imaging supectroscopy
Journal title :
Astroparticle Physics
Record number :
2022929
Link To Document :
بازگشت