Author/Authors :
Moseley، نويسنده , , Greg S.H and Allen MD، نويسنده , , C.A and Benford، نويسنده , , D and Dowell، نويسنده , , C.D and Harper، نويسنده , , D.A. and Phillips، نويسنده , , T.G and Silverberg، نويسنده , , R.F and Staguhn، نويسنده , , J، نويسنده ,
Abstract :
We report on the performance of the SHARC II detector, a 12×32 array of ion implanted Si pop-up bolometers. This 384 element detector array was built as a prototype for the High Angular Resolution Widefield Camera for the Stratospheric Observatory for Infrared Astronomy. We will discuss the design process, the characterization of the detectors, and the performance of the array in the SHARC II instrument. SHARC II is now a facility instrument on the Caltech Submillimeter Observatory, providing background-limited imaging at 350 and 450 μm.
Keywords :
Array detectors , Far-infrared , Submillimeter imaging , Cryogenic detectors