Title of article
Photodetectors for scintillator proportionality measurement
Author/Authors
Moses، نويسنده , , William W. and Choong، نويسنده , , Woon-Seng and Hull، نويسنده , , Giulia and Payne، نويسنده , , Steve and Cherepy، نويسنده , , Nerine and Valentine، نويسنده , , John D.، نويسنده ,
Pages
5
From page
45
To page
49
Abstract
We evaluate photodetectors for use in a Compton Coincidence apparatus designed for measuring scintillator proportionality. There are many requirements placed on the photodetector in these systems, including active area, linearity, and the ability to accurately measure low light levels (which implies high quantum efficiency and high signal-to-noise ratio). Through a combination of measurement and Monte Carlo simulation, we evaluate a number of potential photodetectors, especially photomultiplier tubes and hybrid photodetectors. Of these, we find that the most promising devices available are photomultiplier tubes with high (∼50%) quantum efficiency, although hybrid photodetectors with high quantum efficiency would be preferable.
Keywords
Photomultiplier tubes , Photodetectors , Hybrid photodetectors , Scintillator non-proportionality
Journal title
Astroparticle Physics
Record number
2022983