• Title of article

    Application of a TES micro-calorimeter as high-energy resolution detector for hard X-rays at a synchrotron beam line

  • Author/Authors

    Maurizio، نويسنده , , C. and Pergolesi، نويسنده , , D. Delli Gatti، نويسنده , , F. and d’Acapito، نويسنده , , F. and Razeti، نويسنده , , M. and Balerna، نويسنده , , A. and Mobilio، نويسنده , , S.، نويسنده ,

  • Pages
    3
  • From page
    610
  • To page
    612
  • Abstract
    A superconducting transition edge sensor microcalorimeter was tested at the synchrotron radiation beam-line of the ESRF to collect the X-ray fluorescence spectrum of a Renaissance gold lustre on ceramic. The measured energy resolution of the detector in the range 6–9 keV was 70 eV, i.e. 3 times better than that of the usual solid state detector. The preliminary data collected demonstrate the applicability of these detectors to the energy dispersive X-ray spectroscopy at the synchrotron radiation facilities.
  • Keywords
    Low-temperature detector , Synchrotron radiation , Low-temperature techniques , Cryogenics , Refrigerators , Synchrotron radiation instrumentation
  • Journal title
    Astroparticle Physics
  • Record number

    2023014