Title of article :
Analysis of neutron and photon detection position for the calibration of plastic (BC-420) and liquid (BC-501) scintillators
Author/Authors :
Pozzi، نويسنده , , Sara A. and Mullens، نويسنده , , James A. and Mihalczo، نويسنده , , John T.، نويسنده ,
Abstract :
In this paper, we present the results of the neutron and photon calibration of a BC-420 plastic scintillator and a BC-501 liquid scintillator using a set of reference gamma sources and a Cf-252 source. The position of neutron and photon interactions inside the scintillators as a function of energy is investigated by Monte Carlo simulation.
Keywords :
neutron detection , Photon detection , scintillation detector , Calibration , Monte Carlo
Journal title :
Astroparticle Physics