• Title of article

    Proton therapy beam dosimetry with silicon CMOS image sensors

  • Author/Authors

    D and Sanchez-Crespo، نويسنده , , A and Kerek، نويسنده , , A and Klamra، نويسنده , , W and Molnلr، نويسنده , , J and Norlin، نويسنده , , L.-O and Skatt، نويسنده , , A and Grusell، نويسنده , , E and Novلk، نويسنده , , D and Sipos، نويسنده , , A and Van der Marel، نويسنده , , J and Végh، نويسنده , , J، نويسنده ,

  • Pages
    5
  • From page
    289
  • To page
    293
  • Abstract
    A 16 mm2 CMOS Image Sensor with more than 100 000 pixels and with a standard video output was irradiated with 48, 95 and 180 MeV protons. Proton-induced nuclear reactions in silicon were detected as bright spots or tracks in the images. The angular and energy-dependent response of the detector were studied. The application to proton dosimetry is discussed.
  • Keywords
    CMOS , Radiation dosimetry , proton , Nuclear reaction
  • Journal title
    Astroparticle Physics
  • Record number

    2023556