Title of article :
A radiation-tolerant electronic readout system for portal imaging
Author/Authors :
ضstling، نويسنده , , J and Brahme، نويسنده , , A and Danielsson، نويسنده , , M. and Iacobaeus، نويسنده , , C and Peskov، نويسنده , , V، نويسنده ,
Abstract :
A new electronic portal imaging device, EPID, is under development at the Karolinska Institutet and the Royal Institute of Technology. Due to considerable demands on radiation tolerance in the radiotherapy environment, a dedicated electronic readout system has been designed. The most interesting aspect of the readout system is that it allows to read out ∼1000 pixels in parallel, with all electronics placed outside the radiation beam—making the detector more radiation resistant. In this work we are presenting the function of a small prototype (6×100 pixels) of the electronic readout board that has been tested. Tests were made with continuous X-rays (10–60 keV) and with α particles. The results show that, without using an optimised gas mixture and with an early prototype only, the electronic readout system still works very well.
Keywords :
EPID , GEM , Electronic readout system
Journal title :
Astroparticle Physics