Title of article :
Stabilization of prompt gamma-ray neutron activation analysis (PGNAA) spectra from NaI detectors
Author/Authors :
Metwally، نويسنده , , W.A. and Gardner، نويسنده , , R.P.، نويسنده ,
Pages :
4
From page :
518
To page :
521
Abstract :
NaI detectors are still used frequently in industrial Prompt Gamma-Ray Neutron Activation Analysis applications such as in bulk material analysis. They have the advantages of being efficient for high-energy gamma rays, being relatively rugged, and being able to be used without cooling. When using NaI detectors, and consequently photomultiplier tubes, the quality of the data can drastically deteriorate through gain and zero shifts that result in spectral smearing due to temperature and/or counting rate changes. A new offline approach is presented to stabilize the NaI spectral drift. The approach is not sensitive to the cause of the drift and takes into account the NaI and ADC non-linearities. Peak resolution is improved substantially when this approach is used in the presence of spectral drift.
Keywords :
Spectral drift , Spectrum stabilization , Temperature and counting rate drifts , NaI non-linearity
Journal title :
Astroparticle Physics
Record number :
2023610
Link To Document :