Title of article
Digital pulse processing for planar TlBr detectors
Author/Authors
Nakhostin، نويسنده , , M. and Hitomi، نويسنده , , K. and Ishii، نويسنده , , K. and Kikuchi، نويسنده , , Y.، نويسنده ,
Pages
3
From page
242
To page
244
Abstract
We report on a digital pulse processing algorithm for correction of charge trapping in the planar TlBr detectors. The algorithm is performed on the signals digitized at the preamplifier stage. The algorithm is very simple and is implemented with little computational effort. By using a digitizer with a sampling rate of 250 MSample/s and 8 bit resolution, an energy resolution of 6.5% is achieved at 511 keV with a 0.7 mm thick detector.
Keywords
Thallium bromide , digital signal processing , ?-ray spectrometers
Journal title
Astroparticle Physics
Record number
2023914