Author/Authors :
Eitel، نويسنده , , G. and Block، نويسنده , , M. and Czasch، نويسنده , , A. and Dworschak، نويسنده , , M. and George، نويسنده , , S. and Jagutzki، نويسنده , , O. and Ketelaer، نويسنده , , Benjamin J. and Ketter، نويسنده , , J. and Nagy، نويسنده , , Sz. and Rodrيguez، نويسنده , , D. and Smorra، نويسنده , , C. and Blaum، نويسنده , , K.، نويسنده ,
Abstract :
A commercial, position-sensitive ion detector was used for the first time for the time-of-flight ion-cyclotron resonance detection technique in Penning trap mass spectrometry. In this work, the characteristics of the detector and its implementation in a Penning trap mass spectrometer will be presented. In addition, simulations and experimental studies concerning the observation of ions ejected from a Penning trap are described. This will allow for a precise monitoring of the state of ion motion in the trap.
Keywords :
Beam diagnostics , TOF-ICR , Delay-line detector , Ion Optics , Position-sensitive ion detection , Penning trap mass spectrometry