• Title of article

    Medipix2 as a tool for proton beam characterization

  • Author/Authors

    Bisogni، نويسنده , , M.G. and Cirrone، نويسنده , , G.A.P. and Cuttone، نويسنده , , G. and Del Guerra، نويسنده , , A. and Lojacono، نويسنده , , P. and Piliero، نويسنده , , M.A. and Romano، نويسنده , , F. and Rosso، نويسنده , , V. and Sipala، نويسنده , , V. and Stefanini، نويسنده , , A.، نويسنده ,

  • Pages
    3
  • From page
    48
  • To page
    50
  • Abstract
    Proton therapy is a technique used to deliver a highly accurate and effective dose for the treatment of a variety of tumor diseases. The possibility to have an instrument able to give online information could reduce the time necessary to characterize the proton beam. To this aim we propose a detection system for online proton beam characterization based on the Medipix2 chip. x2 is a detection system based on a single event counter read-out chip, bump-bonded to silicon pixel detector. The read-out chip is a matrix of 256×256 cells, 55×55 μm2 each. To demonstrate the capabilities of Medipix2 as a proton detector, we have used a 62 MeV flux proton beam at the CATANA beam line of the LNS-INFN laboratory. The measurements performed confirmed the good imaging performances of the Medipix2 system also for the characterization of proton beams.
  • Keywords
    Protons , Silicon pixels detectors , Beam quality assurance
  • Journal title
    Astroparticle Physics
  • Record number

    2024149