• Title of article

    Probe prototypes based on TlBr detectors

  • Author/Authors

    Anna and Shorohov، نويسنده , , M. and Gostilo، نويسنده , , V. and Ivanov، نويسنده , , Sanjay V. and Owens، نويسنده , , A.، نويسنده ,

  • Pages
    4
  • From page
    116
  • To page
    119
  • Abstract
    Detectors based on TlBr crystals are ideally suited for the probes, where small size and rapid stopping power are highly desirable attributes. The development results of TlBr probes for medical and nuclear industry applications are presented here. Probes for both applications include TlBr detectors with dimensions of 5×5×2 mm3 and a microelectronics preamplifier. The different construction and technological aspects of probes designed for both application types are discussed. The probes were tested over a wide energy range. We obtained energy resolutions of 3.6, 4.9 and 14.5 keV at energies of 59.6, 122 and 662 keV, respectively, for the developed probe prototypes at room temperature. The observed performances of probe prototypes having TlBr detectors confirmed their promising potential for probe applications.
  • Keywords
    TlBr detectors , Registration efficiency , Probe prototypes
  • Journal title
    Astroparticle Physics
  • Record number

    2024186