Abstract :
We have developed a pixel-type two-dimensional position sensitive Li-glass neutron detector for high rate application, which would be useable for time-of-flight radiography, namely transmission spectroscopy. The pieces of Li-glass were attached directly to a 8×8 channel photo multiplier. We measured its performance at KEK and Hokkaido University. The count rate of this detector was found to be about 8.3 Mcps/cm2 (240 Kcps per pixel). The efficiency of the detector corrected for the effective detection area is 100% over 9 إ and 95% at 3 إ. The difference in the count rate of each pixel was very small, less than 2%.
Keywords :
Li-glass , Pixel type , High count rate , Neutron , 2D detector