• Title of article

    Point dose profile measurements using solid-state detectors in characterization of Computed Tomography systems

  • Author/Authors

    Herrnsdorf، نويسنده , , L. and Bjِrk، نويسنده , , M. and Cederquist، نويسنده , , B. and Mattsson، نويسنده , , C.G. and Thungstrِm، نويسنده , , G. and Frِjdh، نويسنده , , C.، نويسنده ,

  • Pages
    3
  • From page
    223
  • To page
    225
  • Abstract
    Existing solid-state detectors for point dose measurements are typically mounted in a package with a high Z material acting as the back contact causing unsymmetrical response. In the Computed Tomography (CT) dose profile application this is a major drawback. A new silicon detector has been constructed without this disturbing element. The new detector has been mounted in a cylindrical tube and its characteristics are presented in this paper. The conclusion is that this new detector design is a major improvement for correct dose and dose profile measurements in CT systems.
  • Keywords
    Point dose profile measurement , CT QA , CT dose , Solid dose detector
  • Journal title
    Astroparticle Physics
  • Record number

    2024252