Title of article
Low-noise multichannel ASIC for high count rate X-ray diffractometry applications
Author/Authors
Szczygiel، نويسنده , , R. and Grybos، نويسنده , , P. and Maj، نويسنده , , P. and Tsukiyama، نويسنده , , A. and Matsushita، نويسنده , , K. and Taguchi، نويسنده , , T.، نويسنده ,
Pages
4
From page
229
To page
232
Abstract
RG64 is a 64-channel ASIC designed for the silicon strip detector readout and optimized for high count rate X-ray imaging applications. In this paper we report on the test results referring to the RG64 noise level, channel uniformity and the operation with a high rate of input signals. The parameters of the RG64-based diffractometry system are compared with the ones based on the scintillation counter. Diffractometry measurement results with silicon strip detectors of different strip lengths and strip pitch are also presented.
Keywords
x-ray imaging , Silicon strip detector , ASIC
Journal title
Astroparticle Physics
Record number
2024258
Link To Document