Title of article
Optimized readout configuration for PIXE spectrometers based on Silicon Drift Detectors: Architecture and performance
Author/Authors
Alberti، نويسنده , , R. and Grassi، نويسنده , , N. and Guazzoni، نويسنده , , C. and Klatka، نويسنده , , T.، نويسنده ,
Pages
5
From page
458
To page
462
Abstract
An optimized readout configuration based on a charge preamplifier with pulsed-reset has been designed for Silicon Drift Detectors (SDDs) to be used in Particle Induced X-ray Emission (PIXE) measurements. The customized readout electronics is able to manage the large pulses originated by the protons backscattered from the target material that would otherwise cause significant degradation of X-ray spectra and marked increase in dead time. In this way, the excellent performance of SDDs can be exploited in high-quality proton-induced spectroscopy of low- and medium-energy X-rays. This paper describes the designed readout architecture and the performance characterization carried out in a PIXE setup with MeV proton beams.
Keywords
Silicon drift detectors , Backscattered protons , Charge preamplifier , Front-end electronics , PIXE
Journal title
Astroparticle Physics
Record number
2024339
Link To Document