• Title of article

    Absolute 60Co characterization based on gamma–gamma coincident detection by two NaI(Tl) detectors

  • Author/Authors

    Volkovitsky، نويسنده , , Peter and Naudus، نويسنده , , Philip، نويسنده ,

  • Pages
    5
  • From page
    568
  • To page
    572
  • Abstract
    A gamma–gamma coincidence measurement method was used for the characterization of 60Co. A two-dimensional analysis of coincident events allows the careful calculation of the number of counts, when each of the two γ-rays emitted by 60Co is in the photopeak area in each of the NaI(Tl) detectors. The standard formulas that were applied for absolute calibration of 125I (Eldridge–Crowther formulas) were modified for the case of 60Co decay. These modified Eldridge–Crowther formulas were applied to the absolute calibration of 60Co samples measured by the two NaI(Tl) detectors both in coincidence and normal modes. Measurements were performed for three sources of different activities and for two positions of each source between the two NaI(Tl) detectors. Results are in good agreement with the measurements of the same samples done by the 4πβ-γ method.
  • Keywords
    Coincidence gamma-rays , 60Co isotope , NaI(Tl) detector , absolute calibration
  • Journal title
    Astroparticle Physics
  • Record number

    2024378