• Title of article

    Ageing and recovering of glass RPC

  • Author/Authors

    Candela، نويسنده , , A. and De Deo، نويسنده , , M. and Di Giovanni، نويسنده , , A. and Di Marco، نويسنده , , N. and DʹIncecco، نويسنده , , M. and Gustavino، نويسنده , , C. and Redaelli، نويسنده , , N. and Tonazzo، نويسنده , , A. and Trinchero، نويسنده , , G.C.، نويسنده ,

  • Pages
    5
  • From page
    116
  • To page
    120
  • Abstract
    The glass Resistive Plate Chambers efficiency before and after water vapour flushing have been studied. The efficiency has been observed to rapidly drop to the level of 30 % after a few days. After few hours of ammonia flushing, a full recover of the device occurred. Surprisingly, further flushing with water vapour had no effect on the efficiency, suggesting a treatment before the use in an experiment and/or whenever the performance is poor. In this work experimental results on tests at high temperature (up to 55 ∘ C ), high rate (up to 100 Hz / cm 2 ) and ageing (up to 10 mC / cm 2 ) are reported.
  • Keywords
    Rate capability , Glass RPC , Temperature , ageing
  • Journal title
    Astroparticle Physics
  • Record number

    2024657