Title of article :
Effects of radiation damage caused by proton irradiation on Multi-Pixel Photon Counters (MPPCs)
Author/Authors :
Matsumura، نويسنده , , T. and Matsubara، نويسنده , , T. and Hiraiwa، نويسنده , , T. and Horie، نويسنده , , K. and Kuze، نويسنده , , M. and Miyabayashi، نويسنده , , K. and Okamura، نويسنده , , A. and Sawada، نويسنده , , T. and Shimizu، نويسنده , , S. and Shinkawa، نويسنده , , T. and Tsunemi، نويسنده , , T. and Yosoi، نويسنده , , M.، نويسنده ,
Pages :
8
From page :
301
To page :
308
Abstract :
We have investigated the effects caused by proton-induced radiation damage on Multi-Pixel Photon Counter (MPPC), a pixelized photon detector developed by Hamamatsu Photonics. The leakage current of irradiated MPPC samples linearly increases with total irradiated doses due to radiation damage, which is not completely recovered even after a year from the irradiation. No significant change has been observed in the gains at least up to 8.0 Gy ( 9.1 × 10 7 n / mm 2 in 1 MeV neutron equivalent fluence, Φ eq ). The device has completely lost its photon-counting capability due to baseline fluctuations and noise pile-up after 21 Gy irradiation ( 2.4 × 10 8 n / mm 2 in Φ eq ), which might be problematic for some applications, such as ring-imaging Cherenkov detectors. We have found that the pulse-height resolution has been slightly deteriorated after 42 Gy irradiation ( 4.8 × 10 8 n / mm 2 in Φ eq ), where the measured sample has been illuminated with a few hundred photons. This effect should be considered in the case of energy-measurement applications.
Keywords :
MPPC , SiPM , Photon detector , Radiation damage , PPD
Journal title :
Astroparticle Physics
Record number :
2024812
Link To Document :
بازگشت