Title of article :
Preliminary evaluation of a novel energy-resolved photon-counting gamma ray detector
Author/Authors :
Meng، نويسنده , , L.-J. and Tan، نويسنده , , J.W. and Spartiotis، نويسنده , , K. and Schulman، نويسنده , , T.، نويسنده ,
Pages :
7
From page :
548
To page :
554
Abstract :
In this paper, we present the design and preliminary performance evaluation of a novel energy-resolved photon-counting (ERPC) detector for gamma ray imaging applications. The prototype ERPC detector has an active area of 4.4 cm×4.4 cm, which is pixelated into 128×128 square pixels with a pitch size of 350 μm×350 μm. The current detector consists of multiple detector hybrids, each with a CdTe crystal of 1.1 cm×2.2 cm×1 mm, bump-bonded onto a custom-designed application-specific integrated circuit (ASIC). The ERPC ASIC has 2048 readout channels arranged in a 32×64 array. Each channel is equipped with pre- and shaping-amplifiers, a discriminator, peak/hold circuitry and an analog-to-digital converter (ADC) for digitizing the signal amplitude. In order to compensate for the pixel-to-pixel variation, two 8-bit digital-to-analog converters (DACs) are implemented into each channel for tuning the gain and offset. The ERPC detector is designed to offer a high spatial resolution, a wide dynamic range of 12–200 keV and a good energy resolution of 3–4 keV. The hybrid detector configuration provides a flexible detection area that can be easily tailored for different imaging applications. The intrinsic performance of a prototype ERPC detector was evaluated with various gamma ray sources, and the results are presented.
Keywords :
Energy-resolved photon counting , CdTe , Gamma ray imaging
Journal title :
Astroparticle Physics
Record number :
2025553
Link To Document :
بازگشت