Title of article :
Performance evaluation for pinhole collimators of small gamma camera by MTF and NNPS analysis: Monte Carlo simulation study
Author/Authors :
Jeon، نويسنده , , Hosang and Kim، نويسنده , , Hyunduk and Cha، نويسنده , , Bo Kyung and Kim، نويسنده , , Jong Yul and Cho، نويسنده , , Gyuseong and Chung، نويسنده , , Yong Hyun and Yun، نويسنده , , Jong-Il، نويسنده ,
Pages :
4
From page :
93
To page :
96
Abstract :
Presently, the gamma camera system is widely used in various medical diagnostic, industrial and environmental fields. Hence, the quantitative and effective evaluation of its imaging performance is essential for design and quality assurance. The National Electrical Manufacturers Association (NEMA) standards for gamma camera evaluation are insufficient to perform sensitive evaluation. In this study, modulation transfer function (MTF) and normalized noise power spectrum (NNPS) will be suggested to evaluate the performance of small gamma camera with changeable pinhole collimators using Monte Carlo simulation. We simulated the system with a cylinder and a disk source, and seven different pinhole collimators from 1- to 4-mm-diameter pinhole with lead. The MTF and NNPS data were obtained from output images and were compared with full-width at half-maximum (FWHM), sensitivity and differential uniformity. In the result, we found that MTF and NNPS are effective and novel standards to evaluate imaging performance of gamma cameras instead of conventional NEMA standards.
Keywords :
NNPS , Gamma Camera , Sensitivity , FWHM , Uniformity , MTF
Journal title :
Astroparticle Physics
Record number :
2026271
Link To Document :
بازگشت