Title of article :
K-edge subtraction using an energy-resolving position-sensitive detector
Author/Authors :
Gillam، نويسنده , , John E. and Kitcher، نويسنده , , Daniel and Beveridge، نويسنده , , Toby E. and Midgley، نويسنده , , Stewart and Hall، نويسنده , , Chris and Lewis، نويسنده , , Rob A.، نويسنده ,
Pages :
4
From page :
97
To page :
100
Abstract :
Digital Subtraction Angiography is an important technique used to image arterial blood flow using an introduced contrast agent. K-edge subtraction uses contrast-agents with a K-edge feature to acquire the same information but requires knowledge of the X-ray energy and so is usually conducted using synchrotron radiation. However, given a detector that measures position and energy – rather than the currently used integrating devices – it is possible to use white-beam radiation to conduct dynamic K-edge subtraction studies. This study demonstrates an approach to this imaging possibility using analysis of the spectral histogram in each image pixel. K-edge subtraction can be conducted more efficiently using a broad-spectrum X-ray source by processing groups of data either side of the K-edge. Using a simulated data model the possible quality of this new technique is explored under a number of assumed spectral resolutions. The effects on the final image of different methods of image subtraction and pre-processing are also explored.
Keywords :
Position-sensitive detector , K-edge subtraction , Spectral resolution
Journal title :
Astroparticle Physics
Record number :
2026272
Link To Document :