Author/Authors :
Bergamaschi، نويسنده , , A. and Cervellino، نويسنده , , A. and Dinapoli، نويسنده , , R. and Gozzo، نويسنده , , F. and Henrich، نويسنده , , B. and Johnson، نويسنده , , I. and Kraft، نويسنده , , P. and Mozzanica، نويسنده , , A. and Schmitt، نويسنده , , B. and Shi، نويسنده , , X.، نويسنده ,
Abstract :
The Microstrip sYstem for Time rEsolved experimeNts (MYTHEN) detector is a silicon microstrip detector with single photon counting readout developed for powder diffraction experiments at the Swiss Light Source.
dimensional 30 k channels system covering 120° has been installed at the Material Science beamline and is being used for users operation since the end of 2007.
its massively parallel detection of X-rays and fast readout, it is optimized for time-resolved or dose–critical measurements that can be performed in fractions of a second on the whole angular range. The intensity of the synchrotron radiation source can be exploited thanks to the high counting rate capability. Promising results have been obtained also for other synchrotron radiation applications like imaging and pump-and-probe experiments.
tector characteristics will be described in detail and its outstanding performances will be shown.
Keywords :
powder diffraction , Silicon microstrip detector , Photon counting , Detectors for synchrotron radiation