Title of article :
Quantifying the limitations of small animal positron emission tomography
Author/Authors :
Oxley، نويسنده , , D.C. and Boston، نويسنده , , A.J. and Boston، نويسنده , , H.C. and Cooper، نويسنده , , R.J. and Cresswell، نويسنده , , J.R. and Grint، نويسنده , , A.N. and Nolan، نويسنده , , P.J. and Scraggs، نويسنده , , D.P. and Lazarus، نويسنده , , I.H. and Beveridge، نويسنده , , T.E.، نويسنده ,
Pages :
4
From page :
343
To page :
346
Abstract :
The application of position sensitive semiconductor detectors in medical imaging is a field of global research interest. The Monte-Carlo simulation toolkit GEANT4 [http://geant4.web.cern.ch/geant4/] was employed to improve the understanding of detailed γ -ray interactions within the small animal Positron Emission Tomography (PET), high-purity germanium (HPGe) imaging system, SmartPET [A.J. Boston, et al., Oral contribution, ANL, Chicago, USA, 2006]. This system has shown promising results in the field of PET [R.J. Cooper, et al., Nucl. Instr. and Meth. A (2009), accepted for publication] and Compton camera imaging [J.E. Gillam, et al., Nucl. Instr. and Meth. A 579 (2007) 76]. Images for a selection of single and multiple point, line and phantom sources were successfully reconstructed using both a filtered-back-projection (FBP) [A.R. Mather, Ph.D. Thesis, University of Liverpool, 2007] and an iterative reconstruction algorithm [A.R. Mather, Ph.D. Thesis, University of Liverpool, 2007]. Simulated data were exploited as an alternative route to a reconstructed image allowing full quantification of the image distortions introduced in each phase of the data processing. Quantifying the contribution of uncertainty in all system components from detector to reconstruction algorithm allows the areas in need of most attention on the SmartPET project and semiconductor PET to be addressed.
Keywords :
Positron emission tomography , GEANT4 , Imaging , HPGe , Semiconductor
Journal title :
Astroparticle Physics
Record number :
2026337
Link To Document :
بازگشت