Title of article :
Quality of X-ray transmission radiography based on single photon counting pixel device
Author/Authors :
Jakubek، نويسنده , , Jan and Vavrik، نويسنده , , Daniel and Pospisil، نويسنده , , Stanislav and Uher، نويسنده , , Josef، نويسنده ,
Pages :
5
From page :
113
To page :
117
Abstract :
This work investigates the optimization of digital X-ray transmission radiography for the observation of time-dependent processes using a single X-ray photon counting pixel device. The maximization of frame rate of acquired roentgenograms is crucial for this purpose. The influence of the X-ray tube voltage on the exposure time needed to acquire images of sufficient quality was investigated. Image distortions due to non-uniform pixel detection efficiency across the X-ray detector can be successfully eliminated by a “Direct thickness calibration method” proposed here, in spite of X-ray beam hardening due to the variation of inspected body absorption.
Keywords :
digital radiography , Roentgenogram quality , Single photon counting pixel device
Journal title :
Astroparticle Physics
Record number :
2026881
Link To Document :
بازگشت