Author/Authors :
Dubeck?، نويسنده , , F. and Perd’ochov?، نويسنده , , A. and ??epko، نويسنده , , P. and Zat’ko، نويسنده , , B. and Sekerka، نويسنده , , V. and Ne?as، نويسنده , , V. and Sek??ov?، نويسنده , , M. and Hudec، نويسنده , , M. and Bohacek، نويسنده , , P. and Huran، نويسنده , , J.، نويسنده ,
Abstract :
The present work describes a portable digital X-ray scanner based on bulk undoped semi-insulating (SI) GaAs monolithic strip line detectors. The scanner operates in “quantum” imaging mode (“single photon counting”), with potential improvement of the dynamic range in contrast of the observed X-ray images. The “heart” of the scanner (detection unit) is based on SI GaAs strip line detectors. The measured detection efficiency of the SI GaAs detector reached a value of over 60 % (compared to the theoretical one of ∼75 %) for the detection of 60 keV photons at a reverse bias of 200 V. The read-out electronics consists of 20 modules fabricated using a progressive SMD technology with automatic assembly of electronic devices. Signals from counters included in the digital parts of the modules are collected in a PC via a USB port and evaluated by custom developed software allowing X-ray image reconstruction. The collected data were used for the creation of the first X-ray “quantum” images of various test objects using the imaging software developed.
Keywords :
Digital X-ray scanner , SI GaAs strip line detector , X-ray irradiation , digital imaging