Title of article :
Calculation of total efficiencies of extended samples for HPGe detectors
Author/Authors :
Vidmar، نويسنده , , Tim and Likar، نويسنده , , Andrej، نويسنده ,
Pages :
4
From page :
251
To page :
254
Abstract :
Calculation of coincidence summing correction factors in gamma-ray spectrometry can be carried out by making use of full-energy-peak and total efficiencies for a given sample geometry and detector setup. In the case of extended samples, the contribution of gamma-rays that scatter in the sample itself to the total efficiency needs to be taken into account. A method is described for calculating this contribution based on calibration measurements with point sources positioned on the detector housing. The approach is aimed at environmental gamma-ray spectrometry analysis of cylindrical samples on p-type HPGe detectors and is verified against full Monte Carlo calculations of total efficiencies with the GEANT code.
Keywords :
Gamma-ray spectrometry , Total efficiency‎ , Extended samples , Gamma-ray scattering
Journal title :
Astroparticle Physics
Record number :
2027102
Link To Document :