• Title of article

    Design of a single moderator-type neutron spectrometer with enhanced energy resolution in the energy range from a few to 100 keV

  • Author/Authors

    Tanimura، نويسنده , , Y. and Saegusa، نويسنده , , J. and Yoshizawa، نويسنده , , M. and Yoshida، نويسنده , , M.، نويسنده ,

  • Pages
    9
  • From page
    592
  • To page
    600
  • Abstract
    The moderator structure for a neutron spectrometer was optimized with the Monte Carlo code MCNP-4B. The spectrometer consists of a cylindrical moderator and a position-sensitive thermal neutron detector and obtains energy spectra from thermal neutron distribution along its cylindrical axis. The structure of the moderator was improved by using a low hydrogen density material on one end and a high hydrogen density on the other, and inserting a neutron absorber that eliminates thermal neutron diffusion. This design improves the energy resolution of the spectrometer, especially for low-energy neutrons from a few to 100 keV. The designed spectrometer can be applied to the measurement of energy spectra over a neutron energy range from a few keV to 20 MeV.
  • Keywords
    Neutron spectrometer , Moderator , Position-sensitive thermal neutron detector , neutron spectrum , Reference field , MCNP-4B
  • Journal title
    Astroparticle Physics
  • Record number

    2027147