Title of article :
CsI(Tl)/plastic phoswich detector enhanced in low-energy gamma-ray detection
Author/Authors :
K. Yamasoto، نويسنده , , Koutaro and Tsutsumi، نويسنده , , Masahiro and Oishi، نويسنده , , Tetsuya and Yoshizawa، نويسنده , , Michio and Yoshida، نويسنده , , Makoto، نويسنده ,
Pages :
7
From page :
609
To page :
615
Abstract :
A phoswich detector composed of a thin plate CsI(Tl) scintillator and a plastic scintillator (BC-400) has been designed and evaluated in order to improve the sensitivity in the low-energy region of a large-area plastic scintillation detector. This newly designed phoswich detector can be applied to both gross gamma measurement and energy spectrometry for low-energy gamma-ray emitters. Judging by estimations of minimum detectable activity, the lower measurable energy of a large-area plastic scintillation detector can be expanded down to a few tens of keV by adding a thin plate CsI(Tl) scintillator.
Keywords :
Large-area plastic scintillation detector , Phoswich detector , Rise-time discrimination technique , Low-energy region , Thin plate CsI(Tl) scintillator , minimum detectable activity
Journal title :
Astroparticle Physics
Record number :
2027227
Link To Document :
بازگشت