Title of article :
Performance evaluation of a small animal PET scanner. Spatial resolution characterization using 18F and 11C
Author/Authors :
Spinelli، نويسنده , , Antonello E. and D’Ambrosio، نويسنده , , Daniela and Pettinato، نويسنده , , Cinzia and Trespidi، نويسنده , , Silvia and Nanni، نويسنده , , Cristina and Ambrosini، نويسنده , , Valentina and Baldazzi، نويسنده , , Giuseppe and Bergamini، نويسنده , , Carlo and Marengo، نويسنده , , Mario، نويسنده ,
Pages :
4
From page :
215
To page :
218
Abstract :
The performance of a depth-of-interaction preclinical small animal PET scanner (GE eXplore Vista) has been investigated. To this purpose several parameters such as spatial resolution, sensitivity, scatter fraction and noise equivalent counts were measured. Because of the different mean positron range of 18F and 11C attention was focused on measuring the effect of such difference on the scanner spatial resolution. The spatial resolution was measured using a point, a line source and a high resolution phantom. Scatter fraction and noise equivalent counts were measured using mouse and rat size objects. Results show that the eXplore Vista provides a spatial resolution of 1.48 mm and sensitivity up to 6%. In vivo images of small mice were also presented.
Keywords :
Small animal PET scanner , Scatter fraction , Noise equivelent counts , Spatial resolution
Journal title :
Astroparticle Physics
Record number :
2027319
Link To Document :
بازگشت