Author/Authors :
Scafè، نويسنده , , Raffaele and Pani، نويسنده , , Roberto and Pellegrini، نويسنده , , Rosanna and Iurlaro، نويسنده , , Giorgia and Montani، نويسنده , , Livia and Nerina Cinti، نويسنده , , Maria، نويسنده ,
Abstract :
In this work, we report and compare the results of γ-ray spectroscopy measurements performed using a LaBr3:Ce crystal coupled to a photomultiplier tube with modified base and to a Si-avalanche photodiode readout by standard electronics. Energy resolution values have been obtained in the range from 59.5 keV (241Am) to 1332 keV (60Co). Good linearity in energy response has been found. Energy resolution values obtained with avalanche photodiode readout are underestimated due to the small active area of the presently available photosensor with respect to crystal dimension.
Keywords :
LaBr3:Ce , scintillator , Nuclear medicine , ?-Ray spectroscopy