Title of article
Si-APD readout for LaBr3:Ce scintillator
Author/Authors
Scafè، نويسنده , , Raffaele and Pani، نويسنده , , Roberto and Pellegrini، نويسنده , , Rosanna and Iurlaro، نويسنده , , Giorgia and Montani، نويسنده , , Livia and Nerina Cinti، نويسنده , , Maria، نويسنده ,
Pages
3
From page
355
To page
357
Abstract
In this work, we report and compare the results of γ-ray spectroscopy measurements performed using a LaBr3:Ce crystal coupled to a photomultiplier tube with modified base and to a Si-avalanche photodiode readout by standard electronics. Energy resolution values have been obtained in the range from 59.5 keV (241Am) to 1332 keV (60Co). Good linearity in energy response has been found. Energy resolution values obtained with avalanche photodiode readout are underestimated due to the small active area of the presently available photosensor with respect to crystal dimension.
Keywords
LaBr3:Ce , scintillator , Nuclear medicine , ?-Ray spectroscopy