Title of article :
An X-ray scanner prototype based on a novel hybrid gaseous detector
Author/Authors :
Iacobaeus، نويسنده , , C. and Francke، نويسنده , , T. and Lund-Jensen، نويسنده , , B and Peskov، نويسنده , , V.، نويسنده ,
Pages :
4
From page :
61
To page :
64
Abstract :
We have developed a prototype of a new type of hybrid X-ray detector. It contains a thin wall (few μm) edge- illuminated lead glass capillary plate (acting as a converter of X-rays photons to primary electrons) combined with a microgap parallel-plate avalanche chamber operating in various gas mixtures at 1 atm. The operation of these converters was studied in a wide range of X-ray energies (from 6 to 60 keV) at incident angles varying from 0° to 90°. The detection efficiency, depending on the geometry, photonʹs energy, incident angle and the mode of operation, was between a few and ∼40%. The position resolution achieved was ∼50 μm in digital form and was practically independent of the photonʹs energy or gas mixture. veloped detector may open new possibilities for medical imaging, for example in mammography, portal imaging, radiography (including security devices), crystallography and many other applications.
Keywords :
mammography , Capillary plates , Gaseous detector