Author/Authors :
Juillard، نويسنده , , A. and Marnieros، نويسنده , , S. and Dolgorouky، نويسنده , , Y. and Bergé، نويسنده , , L. and Collin، نويسنده , , S. and Fiorucci، نويسنده , , S. and Lalu، نويسنده , , F. and Dumoulin، نويسنده , , L.، نويسنده ,
Abstract :
The actual limitation of Ge ionization heat cryogenic detectors for direct WIMP detection such as EDELWEISS arises from incomplete charge collection for near-surface events. We present results on Ge/NbSi detectors that are fitted with segmented electrodes and two NbSi Anderson insulator thermometric layers. Three such bolometers were studied in the low-background cryostat of the EDELWEISS collaboration in the LSM: analysis of the athermal signals allows us to identify and reject events occurring in the first millimeter under the electrodes.
Keywords :
Dark matter , Surface events , bolometers , WIMPs , Anderson insulators