• Title of article

    The dynamic range of ultra-high-resolution cryogenic gamma-ray spectrometers

  • Author/Authors

    Ali، نويسنده , , Shafinaz and Terracol، نويسنده , , Stephane F. and Drury، نويسنده , , Owen B. and Friedrich، نويسنده , , Stephan، نويسنده ,

  • Pages
    3
  • From page
    542
  • To page
    544
  • Abstract
    We are developing high-resolution cryogenic gamma-ray spectrometers for nuclear science and non-proliferation applications. The gamma-ray detectors are composed of a bulk superconducting Sn foil absorber attached to a multilayer Mo/Cu transition-edge sensor (TES). The energy resolution of a detector with a 1×1×0.25 mm3 Sn absorber is 50–90 eV FWHM for γ-rays up to 100 keV, and it decreases for larger absorbers. Here, we present the detector performance for different absorber volumes, and discuss the trade-offs between energy resolution and dynamic range.
  • Keywords
    Microcalorimeter , Low-temperature detector , Gamma-ray spectroscopy , dynamic range
  • Journal title
    Astroparticle Physics
  • Record number