Author/Authors :
Bazzarri، نويسنده , , S. and Cicoli، نويسنده , , G. and De Felice، نويسنده , , P. and Rossi، نويسنده , , G. and Sedda، نويسنده , , F.، نويسنده ,
Abstract :
Experimental tests are described to check solid 241Am radioactive sources used to monitor operation of liquid xenon detectors. In particular, radioactive leakage was tested in extreme temperature conditions following immersion in liquid nitrogen for different time periods. No radioactivity loss was detected from the sources considered. The paper describes the source characteristics, test methods and results obtained.