Title of article :
Monte Carlo optimization of sample dimensions of an 241Am–Be source-based PGNAA setup for water rejects analysis
Author/Authors :
Idiri، نويسنده , , Z. and Mazrou، نويسنده , , H. and Beddek، نويسنده , , S. and Amokrane، نويسنده , , A. and Azbouche، نويسنده , , A.، نويسنده ,
Abstract :
The present paper describes the optimization of sample dimensions of a 241Am–Be neutron source-based Prompt gamma neutron activation analysis (PGNAA) setup devoted for in situ environmental water rejects analysis. The optimal dimensions have been achieved following extensive Monte Carlo neutron flux calculations using MCNP5 computer code. A validation process has been performed for the proposed preliminary setup with measurements of thermal neutron flux by activation technique of indium foils, bare and with cadmium covered sheet. Sensitive calculations were subsequently performed to simulate real conditions of in situ analysis by determining thermal neutron flux perturbations in samples according to chlorine and organic matter concentrations changes. The desired optimal sample dimensions were finally achieved once established constraints regarding neutron damage to semi-conductor gamma detector, pulse pile-up, dead time and radiation hazards were fully met.
Keywords :
PGNAA technique , Monte Carlo calculations , MCNP5 , Semi-conductor gamma detector , Optimal sample dimensions , activation foil technique
Journal title :
Astroparticle Physics