• Title of article

    MCNP simulation and spectrum unfolding for an NaI monitor of radioactivity in aquatic systems

  • Author/Authors

    Mertens، نويسنده , , C. and De Lellis، نويسنده , , C. and Van Put، نويسنده , , P. and Tondeur، نويسنده , , F.، نويسنده ,

  • Pages
    5
  • From page
    118
  • To page
    122
  • Abstract
    Spectrum unfolding can be implemented when the angular distribution of the radiation fluence is known, e.g. when the source–detector geometry is fixed. In the case of the system developed by Institut National des Radio-Elements (IRE) for monitoring radioactivity in aquatic systems, the source is an infinite homogenous medium surrounding the monitor, with constant density. Its response function for gamma rays is calculated with MCNP4.C2. The simulation model is validated with reference measurements of standard sources diluted in a drum. Results of unfolding with UMG 3.3 are presented for this situation, and for a measurement in sea water.
  • Keywords
    Spectrum unfolding , Gamma-ray monitor , SIMULATION , MCNP
  • Journal title
    Astroparticle Physics
  • Record number