Evaluation of a multi-guard ring (MGR) structure diode as diagnostic X-ray dosimeter
Author/Authors :
Camargo، نويسنده , , F. and Khoury، نويسنده , , H.J. and Nascimento، نويسنده , , C.R. and Asfora، نويسنده , , V.K. and Bueno، نويسنده , , C.C.، نويسنده ,
Abstract :
In this paper, we describe the results obtained for the evaluation of a multi-guard ring (MGR) structure diode as diagnostic X-ray dosimeter. This device was developed in the framework of R&D programs for the future CMS experiment at the Large Hadron Collider (LHC) with high radiation hardness to fulfill the requirements from this accelerator environment. In order to use the MGR diode as a dosimeter, it was connected in the photovoltaic mode to the input of an integrating electrometer and positioned at the center of an X-ray beam, beside a previously calibrated ionization chamber. The dependence of the diode response on the X-ray beam doses was evaluated for 35–90 kV X-ray generator bias supply, with doses in the range of 50 μGy–5 mGy. The good linearity of the dose–response curve obtained showed the MGR diode dosimeter to be a reliable alternative method for diagnostic X-ray dosimetry.
Keywords :
x-ray dosimetry , Silicon detector , Radiation Protection , Rad-hard silicon detector , Photon dosimetry