Title of article :
Design and preliminary Monte Carlo calculations of an active Compton-suppressed LaBr3(Ce) detector system for TRU assay in remote-handled wastes
Author/Authors :
Kulisek، نويسنده , , J.A. and Hartwell، نويسنده , , J.K. and McIlwain، نويسنده , , M.E. and Gardner، نويسنده , , R.P.، نويسنده ,
Pages :
4
From page :
226
To page :
229
Abstract :
Recent studies indicate LaBr3(Ce) scintillating detectors have desirable attributes, such as room temperature operability and excellent energy resolution, which may make them viable alternatives as primary detectors (PD) in a Compton suppression spectrometer (CSS) used for remote-handled transuranic (RH-TRU) waste assay. A CSS with a LaBr3(Ce) PD has been designed and its expected performance evaluated using Monte Carlo analysis. These results indicate that this detector will have a relatively high Compton-suppression capability, with greater suppression ability for large angle-scattered photons in the PD.
Keywords :
Compton suppression , MCNP , Scintillation detectors , Gamma-ray spectrometry
Journal title :
Astroparticle Physics
Record number :
2029975
Link To Document :
بازگشت