Title of article :
Monte Carlo and least-squares methods applied in unfolding of X-ray spectra measured with cadmium telluride detectors
Author/Authors :
Moralles، نويسنده , , M. and Bonifلcio، نويسنده , , D.A.B. and Bottaro، نويسنده , , M. and Pereira، نويسنده , , M.A.G.، نويسنده ,
Pages :
4
From page :
270
To page :
273
Abstract :
Spectra of calibration sources and X-ray beams were measured with a cadmium telluride (CdTe) detector. The response function of the detector was simulated using the GEANT4 Monte Carlo toolkit. Trapping of charge carriers were taken into account using the Hecht equation in the active zone of the CdTe crystal associated with a continuous function to produce drop of charge collection efficiency near the metallic contacts and borders. The rise time discrimination is approximated by a cut in the depth of the interaction relative to cathode and corrections that depend on the pulse amplitude. The least-squares method with truncation was employed to unfold X-ray spectra typically used in medical diagnostics and the results were compared with reference data.
Keywords :
CdTe detector , Monte Carlo , diagnostic x-ray , GEANT4
Journal title :
Astroparticle Physics
Record number :
2029986
Link To Document :
بازگشت