• Title of article

    Photon and fast neutron dosimetry using aluminium oxide thermoluminescence dosemeters in a pool-type research reactor

  • Author/Authors

    Santos، نويسنده , , J.P. and Marques، نويسنده , , J.G. and Fernandes، نويسنده , , A.C. and Osvay، نويسنده , , M.، نويسنده ,

  • Pages
    4
  • From page
    310
  • To page
    313
  • Abstract
    Al2O3:Mg,Y thermoluminescence (TL) dosemeters were used to measure photon and fast neutron doses in the mixed radiation field of the Portuguese Research Reactor. The dosemeters were irradiated in core positions under a photon dose rate of the order of 104 Gy/h and a fast neutron flux in the range of 109–1011 n/cm2/s. In order to evaluate the ability of the TL dosemeters for mixed field dosimetry at the research reactor, the measurements were compared with results obtained via conventional methods. The agreement between the different methods is better than 13% for the determination of photon doses and within 5% for the determination of neutron fluxes in mixed fields.
  • Keywords
    THERMOLUMINESCENCE , dosimetry , Mixed field , D-2 dosemeters , Aluminium oxide
  • Journal title
    Astroparticle Physics
  • Record number

    2029997