Author/Authors :
Fujita، نويسنده , , K. and Takahashi، نويسنده , , H. and Siritiprussamee، نويسنده , , P. and Niko، نويسنده , , H. and Nishi، نويسنده , , K. and Takada، نويسنده , , Y. and Oku، نويسنده , , T. and Suzuki، نويسنده , , J. and Ino، نويسنده , , T. and Shimizu، نويسنده , , H.M and Furusaka، نويسنده , , M.، نويسنده ,
Abstract :
High-resolution neutron imaging detectors are required for future high-intensity neutron sources such as the J-PARC project in Japan. We are developing a high-resolution gaseous neutron detector by using the microstrip gas chamber (MSGC) with floating-pad readout method. In this paper, we describe the neutron detector performance obtained with this MSGC and the readout method. For a 3 cm×3 cm MSGC filled with 5 bar CF4 and 0.5 bar 3He, we could achieve a measured position resolution of 0.6 mm (FWHM) including the beam size and the parallax error.
Keywords :
M-MSGC , Floating pad , 2D imaging , high resolution , neutron detector