Author/Authors :
Nagayoshi، نويسنده , , Tsutomu and Doke، نويسنده , , Tadayoshi and Fujita، نويسنده , , Yasunobu and Hattori، نويسنده , , Kaori and Ishida، نويسنده , , Koji and Kikuchi، نويسنده , , Jun and Kitamura، نويسنده , , Hisashi and Komiyama، نويسنده , , Tatsuto and Kubo، نويسنده , , Hidetoshi and Matsumoto، نويسنده , , Haruhisa and Miuchi، نويسنده , , Kentaro and Nishimura، نويسنده , , Hironobu and Saito، نويسنده ,
Abstract :
Beam tests were performed for a Micro Pixel Chamber ( μ -PIC) with a detection volume of 10 × 10 × 10 cm 3 to investigate the response to heavy ions. The three-dimensional tracks of carbon, silicon, and iron beams were successfully observed and their track lengths were measured. Additionally, Linear Energy Transfer (LET) distributions of each ion were obtained, and the mean LET values were consistent with the theoretical calculation of mass stopping power within an error of ∼ 10 % . This detector is a candidate for an ideal dosimeter in space.
Keywords :
LET , Position sensitive , TEPC , Gaseous detector , Heavy ions , Micro-pattern detector