Title of article :
Isochronism of the ACCEL 250 MeV medical proton cyclotron
Author/Authors :
Baumgarten، نويسنده , , C. and Geisler، نويسنده , , A. and Klein، نويسنده , , U. and Krischel، نويسنده , , D. and Rِcken، نويسنده , , H. and Schillo، نويسنده , , M. and Stephani، نويسنده , , T. and Timmer، نويسنده , , J.H.، نويسنده ,
Pages :
5
From page :
10
To page :
14
Abstract :
Beam phase measurements performed during the commissioning of the ACCEL 250 MeV medical proton cyclotrons are presented and compared to results of orbit tracking computations based on high accuracy magnetic field map measurements. The agreement between measured and intended beam phase behavior was found to be excellent.
Keywords :
cyclotron , Isochronism , Beam phase