Title of article
Detection of fast neutrons with the Medipix-2 pixel detector
Author/Authors
Uher، نويسنده , , J. and Jakubek، نويسنده , , J. and Koster، نويسنده , , U. and Lebel، نويسنده , , C. and Leroy، نويسنده , , C. and Pospisil، نويسنده , , S. and Skoda، نويسنده , , R. and Vykydal، نويسنده , , Z.، نويسنده ,
Pages
4
From page
71
To page
74
Abstract
Neutron radiography using thermal and fast neutrons is becoming increasingly important for scientific, technical, security and other applications. In the past, the Medipix-2 imaging detector was successfully adapted for thermal neutron imaging by our group by adding a 6LiF neutron converter covering the surface of the active part of the detector. Recently, the Medipix-2 detector was also adapted and used for fast neutron imaging. Fast neutrons are detected through proton recoil from a 1 mm thick polyethylene layer placed on the detector surface. Basic detection and imaging properties of the detector were calculated, simulated and measured. The measurements were done using an AmBe neutron source, as well as external neutron beams of the Sparrow reactor at the Czech Technical University in Prague and of the high-flux reactor at the Institut Laue-Langevin (ILL) in Grenoble. The detector provided a reasonable signal-to-background ratio of about eight at the neutron beam of ILLʹs Neutrograph neutron radiography and tomography station. The estimated spatial resolution was at a level of 100 μm.
Keywords
Pixel detectors , neutron detection , Neutron imaging , fast neutrons
Journal title
Astroparticle Physics
Record number
2031971
Link To Document