Title of article
Pixel detectors for imaging with heavy charged particles
Author/Authors
Jakubek، نويسنده , , Jan and Cejnarova، نويسنده , , Andrea and Holy، نويسنده , , Tomas and Pospisil، نويسنده , , Stanislav and Uher، نويسنده , , Josef and Vykydal، نويسنده , , Zdenek، نويسنده ,
Pages
4
From page
155
To page
158
Abstract
State-of-the-art single quantum counting pixel detectors offer a large potential for different imaging applications. The TimePix pixel device can provide information about position and energy of the detected radiation allowing radiography with charged particles. Heavy charged particles of known initial energy lose their energy partially by going through a specimen material. If the resulting energies of particles passing the specimen are measured, then specimen structure can be revealed. This article shows experimental results of this technique acquired with alpha particles and the TimePix detector. The spatial resolution in detector plane depends on particle energy and can reach submicrometer level. The specimen thickness can be determined with precision up to 320 nm for organic materials if energy loss of individual alpha particle is measured.
Keywords
Spectroscopy , Radiography , Pixel detector
Journal title
Astroparticle Physics
Record number
2032023