• Title of article

    Investigation of pixelated TlBr gamma-ray spectrometers with the depth-sensing technique

  • Author/Authors

    Hitomi، نويسنده , , Keitaro and Onodera، نويسنده , , Toshiyuki and Shoji، نويسنده , , Tadayoshi and He، نويسنده , , Zhong، نويسنده ,

  • Pages
    3
  • From page
    276
  • To page
    278
  • Abstract
    Pixelated thallium bromide (TlBr) detectors with a thickness of 4.2 mm have been fabricated and investigated with the depth-sensing technique. The energy resolution of the device was improved from 1.95% FWHM to 1.25% FWHM at 662 keV by implementation of the depth correction. The device exhibited an energy resolution of 1.03% FWHM at 662 keV by selecting gamma-ray interaction events occurring underneath the cathode.
  • Keywords
    Thallium bromide , Semiconductor detectors , Position-sensitive detectors
  • Journal title
    Astroparticle Physics
  • Record number

    2032102