Monte Carlo modelling of germanium crystals that are tilted and have rounded front edges
Author/Authors :
Gasparro، نويسنده , , Joel and Hult، نويسنده , , Mikael and Johnston، نويسنده , , Peter N. and Tagziria، نويسنده , , Hamid، نويسنده ,
Abstract :
Gamma-ray detection efficiencies and cascade summing effects in germanium detectors are often calculated using Monte Carlo codes based on a computer model of the detection system. Such a model can never fully replicate reality and it is important to understand how various parameters affect the results. This work concentrates on quantifying two issues, namely (i) the effect of having a Ge-crystal that is tilted inside the cryostat and (ii) the effect of having a model of a Ge-crystal with rounded edges (bulletization).
fect of the tilting is very small (in the order of per mille) when the tilting angles are within a realistic range. The effect of the rounded edges is, however, relatively large (5–10% or higher) particularly for gamma-ray energies below 100 keV.
Keywords :
Monte Carlo simulations , Gamma-ray spectrometry , Rounded edge , Tilted crystal , Bulletization , Low-level