Title of article :
Energy calibration of gamma spectra in plastic scintillators using Compton kinematics
Author/Authors :
Siciliano، نويسنده , , E.R. and Ely، نويسنده , , J.H. and Kouzes، نويسنده , , R.T. and Schweppe، نويسنده , , J.E. and Strachan، نويسنده , , D.M. and Yokuda، نويسنده , , S.T.، نويسنده ,
Pages :
12
From page :
232
To page :
243
Abstract :
This paper describes a simple and practicable method for assigning energy values to gamma-ray pulse-height distributions measured with polyvinyl toluene (PVT) based detectors. It is based upon the characteristic shape of simulated spectra in the region of maximum energy deposition resulting from a single Compton scattering. The validity of this method is demonstrated by applying it to a set of measured NaI(Tl) spectra, and comparing those results to the standard photopeak method of calibrating the same spectra. The method is then applied to a set of measured PVT spectra. It can now be applied to calibration of fielded PVT detectors using two selected gamma-ray sources, which was the initial motivation for this study.
Keywords :
gamma-ray detection , Portal monitor , Radiation detection , homeland security , Detection of illicit materials , Monte Carlo modeling , detector calibration , Border security
Journal title :
Astroparticle Physics
Record number :
2032630
Link To Document :