Author/Authors :
King III، نويسنده , , George S. and Avignone III، نويسنده , , Frank T. and Cox، نويسنده , , Christopher E. and Hossbach، نويسنده , , Todd W. and Jennings، نويسنده , , Wayne and Reeves، نويسنده , , James H.، نويسنده ,
Abstract :
A p-type semi-coaxial HPGe detector has been segmented by cutting, with a diamond saw, and etching four circumferential grooves through the Li-diffused dead layer. The degree of segmentation was tested using a well-collimated low-energy gamma-ray source. An analysis cut that rejected events depositing energy in more than one segment was applied to an energy interval of 2038±5 keV, the region of interest (Qββ) for 76Ge 0νββ decay experiments. This segmentation cut resulted in a reduction of the Compton continuum of 59%.